Key: test item, instrument name, total nonvolatile migration, gravimeter, moisture permeameter
Product

Particle Protection Effect Tester
- Description
- Applications
- Standard
- Features
- Parameter
- Configuration
- Video
- PDF Download
-
- Commodity name: Particle Protection Effect Tester
PPP-2020
Function
It is used to test the protective effect and respiratory resistance of respirators.
Principle
Wear the mask to be inspected on the headform,and place the standard head form in the test chamber was filled with Salt aerosol or oily aerosol with standard particle size and concentration. A sinusoidal airflow was used to simulate breathing. The concentration of salt aerosol or oil aerosol before and after the mask was detected by high-frequency sampling with a photometer, and the protective effect of the mask on the particulate matter was calculated.
Standards
GB/T6165-2008 、GB/2626-2006 、GB/T 32610-2016Characteristics
● Using a built-in high-precision laser scattering photometer, guarantee accurate measurement of aerosol concentration.
● Respirator protection effect test and respiratory resistance detection test can be performed at the same time.
● The standard salt and oil aerosol generator can automatically generate continuous aerosol with specific particle size and concentration;
● A built-in precision flow controller, ensure more accurate flow control.
● Automatic control of the test process, real-time measurement of aerosol concentration, resistance, flow, automatic calculation of filtration efficiency.
● Provide a function to save, query, and print test data in the database.
● 10.1-inch color high-resolution touch screen, with an intuitive interface, easy to use.
● Leakproof design of test chamber, safer when open and close glass door.
SpecificationName
Parameter
Respirator protection effect
0.001% ~ 99.999%
Error
0.5 %
Range of photometer
0.001 ~ 300 mg/m3
Precision of photometer
1%
sampling flow
1~2L/min
Aerosol concentration
25mg/m3 ±10%
Flow control
0~100 L/min
Respiratory rate
20 /min
Respiratory flow
30 L/min
Simulated respiratory airflow
Sine curve
Test decision mode
Timing(1h ) or pressure difference(150 Pa)
The median diameter of aerosol mass
Salt particles (0.6 um), oil particles(0.3 um)
Respiratory resistance test flow
85 ±0.1L/min
The detection range of respiratory resistance
0~1000 Pa, precision:1Pa
Size
1300(L)x650(B)x1900(H)mm
Net weight
280Kg
Power supply
AC 220V50Hz 1500W
Configuration
Principal machine(1), salt aerosol generator(1),oil aerosol generator(1), headform(1),professional computer software(1), pipe fittings(1), power cord(1), communication data line(1).
Self-SupplyAir compressor.
Keywords:
Previous
Related Products
Product Consultation
Contact Us
Guangzhou Labstone Technology Co., Ltd
Add: 6 / F, building A1, No. 2-6, Jinbao Road, Baiyun Hubao street, Baiyun District, Guangzhou, Guangdong
Technical:020-86406337
Switchboard:020-86406230
Email:LabStone@vip.163.com
Message
fOLLOW